E-katalog skupiny BLOCK®
Alpha port - RTP (Rapid Transfer Port) 510.53/A
RTP (Rapid Transfer Port) is a device that is used to securely transfer material from the chamber to elsewhere, while preventing contamination of the product or the environment. RTP consists of two basic parts. Alpha port is fixed on the isolator chamber; Beta port, which is connected to Alpha port, is used as a container for material transfer. RTP is based on a double-door principle; the port into the container may not be opened until the Alpha and Beta ports are connected. RTP is often used for material output and transport or for the removal of solid waste from the chambers. The system comes in five standard sizes 105, 190, 270, 350 and 460 mm. The RTP system is manufactured from materials that are resistant to multiple sterilization (autoclave, ethylene oxide, gamma sterilization, H2O2 sterilization, etc.).
The RTP system features the following safety systems to protect against accidental detachment:
- The Alpha door cannot be opened if the Beta container is incorrectly connected or placed.
- When the Alpha door is opened, the Beta container cannot be disconnected.
TECHNICAL DATA
Type designation of port |
---|
Type designation of port |
Selecting dimensional range |
---|
Alpha Ø105 |
Alpha Ø190 |
Alpha Ø270 |
Alpha Ø350 |
In this category, you can select the door size. |